FDH-0601 Straight-run lubricant oxidation stability analyzer
Scope
Straight run lubricant oxidation stability analyzer based SH / T0185 design and manufacturing requirements, the two oxidation mounted sample tube into the heating temperature of 120 ° bath in oxygen 48H, volatile acids by the flow of oxygen blowing, water is absorbed. We need to determine the amount of volatile acid value, acid value and oil soluble precipitate after the end of oxidation. Oxidation stability is represented by the percentage of the total oxidation products.
Features
1, straight run lubricant oxidation stability analyzer using liquid crystal display, English operation interface
2, straight-run lubricating oil oxidation stability tester using SCM automatic control experiment without human duty
3, straight-run changes in lubricant oxidation stability analyzer assay of various parameters
4, straight run lubricant oxidation stability analyzer has a self-diagnostic function, suggesting operating functions
5, straight-run lubricating oil oxidation stability tester metal bath heating, environmental safety
Technical Parameters
1. Applicable standards: SH / T 0185
2, the heating method: heating tube
3, the control mode: Auto Control
4. Timing mode: SCM automatically record
5, results processing: automatically records storage Results
6, pressure display: LCD display real-time curve
7, Power supply: AC220V / 50HZ
8, Total Power: 2.5KW