FDQ-1071 Salt content analyzer
Scope
Salt content analyzer by computer control and measurement electrolysis, and data processing, computer electrolytic current integral obtained supplementary consumption Ag + power required, the salt content of the sample determined according to Faraday's law. Applies to crude oil, residual oil, oil products and various other industrial water emissions inorganic salt content in water samples and determination of chloride ions, it is an ideal instrument current electric desalting process control salt content analysis.
Features
1, it uses computer-controlled, INDOWS operating platform, English interface
2, it uses automatically determine the end result, without human duty
3, automatic data processing, preservation, and print reports
Technical Parameters
1, measuring range: salt content: 0.2 mg / l ~ 10000 mg / l
2. Sensitivity: salt content: 0.1 mg / l; inorganic Cl-: 0.1mg / l
3, Repeatability:
Salt content ≤2mgNaCl / l absolute error: ≤ ± 0.2mgNaCl / l
Salt content> 2mgNaCl / l relative error: ≤ ± 5%
4, measuring time: